Abstract :
The melting plateaus of W-γW2C eutectic point in graphite and tungsten crucibles were investigated by means of radiation thermometry for the first time. W-cell fixed point showed good repeatability of the melting temperature: it can provide the means of realizing high-temperature fixed points for on site calibration of radiation thermometer in an induction furnace for the measurement of melting temperatures of nuclear fuels. The temperature value of W-γW2C eutectic point will be determined precisely by use of graphite cell. An effect of temperature uniformity of an induction furnace on the melting plateau shape is also determined by use of Al2O3 fixed point.
Keywords :
calibration; fission reactor fuel; graphite; melting point; pyrometers; temperature measurement; thermometers; tungsten; W-γW2C eutectic point; high temperature tungsten cell; induction furnace; melting temperature; nuclear fuel; pyrometer calibration; radiation thermometry; temperature measurement; Aluminum oxide; Furnaces; Materials; Temperature distribution; Temperature measurement; Tungsten; eutectic; fixed point; high temperature; nuclear fuel; radiation thermometer; temperature standard;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
SICE Annual Conference 2010, Proceedings of
Conference_Location :
Taipei
Print_ISBN :
978-1-4244-7642-8
Type :
conf
Filename :
5602269
Link To Document :
بازگشت