Abstract
The possibility to track the threshold voltage shifts caused by ionizing radiation in MOS dosimeters in a Zero Temperature Coefficient reference current is studied. 140 nm gate oxide p-channel MOS transistors were irradiated with a high dose rate 60Co γ-source to characterize their sensitivity to radiation. Results show that the error introduced by temperature can be minimized in one order of magnitude if the ZTC reference current is applied, but that the error increases as the device is irradiated.
Keywords
MOSFET; dosimeters; measurement errors; temperature measurement; γ-source; MOS dosimeters ionizing radiation; ZTC reference current; gate oxide p-channel MOS transistor; low dose radiation measurement; size 140 nm; temperature error minimization; threshold voltage shift; zero temperature coefficient; Current measurement; Logic gates; Radiation effects; Sensitivity; Temperature; Temperature measurement; Temperature sensors; Dosimetry; MOSFETs; Radiation monitoring; Temperature;
fLanguage
English
Publisher
ieee
Conference_Titel
Argentine School of Micro-Nanoelectronics Technology and Applications (EAMTA), 2010
Conference_Location
Montevideo
Print_ISBN
978-1-4244-6747-1
Electronic_ISBN
978-987-1620-14-2
Type
conf
Filename
5606372
Link To Document