Title :
Reduction arithmetic for power line interference from ECG based on estimating sinusoidal parameters
Author :
Hu, Xiao ; Xiao, Zhong ; Liu, Changhong
Author_Institution :
Dept. of Electron. & Inf. Eng., Guangzhou Univ., Guangzhou, China
Abstract :
In this paper, a novel algorithm was proposed to subtract power line interference (PLI) from contaminated ECG signal. Firstly, some PLI values were calculated from ECG signal in linear segment. Secondly, using nonlinear regression and least squares estimation, sinusoidal parameters such as frequency and phase were estimated from the calculated PLI values. Thirdly, the PLI values in the nonlinear segment closest to the linear segment were determined by the sinusoid function with the estimated parameters. Lastly, PLI was subtracted from the contaminated ECG signal in nonlinear segment. The method´s performance was tested in ECG databases in www.physionet.org. The experiment results show that the proposed method can more effectively remove PLI from ECG signal than other algorithms even when PLI frequency fluctuates, and do not distort the shape of ECG signal.
Keywords :
electrocardiography; electromagnetic interference; least squares approximations; medical signal processing; regression analysis; signal denoising; ECG signal shape; PLI frequency; contaminated ECG signal; least squares estimation; nonlinear regression; power line interference subtraction; reduction arithmetic; sinusoidal parameter estimation; Damping; Digital filters; Distortion; Electrocardiography; Interference; Low pass filters; Silicon; electrocardiogram; levenberg-marquardt algorithm; nonlinear regression; power line interference;
Conference_Titel :
Biomedical Engineering and Informatics (BMEI), 2010 3rd International Conference on
Conference_Location :
Yantai
Print_ISBN :
978-1-4244-6495-1
DOI :
10.1109/BMEI.2010.5640006