DocumentCode :
53589
Title :
An Adaptive Wearable Parallel Robot for the Treatment of Ankle Injuries
Author :
Jamwal, P.K. ; Xie, Sheng Q. ; Hussain, Shiraz ; Parsons, John G.
Author_Institution :
Rajasthan Tech. Univ., Kota, India
Volume :
19
Issue :
1
fYear :
2014
fDate :
Feb. 2014
Firstpage :
64
Lastpage :
75
Abstract :
This paper presents the development of a novel adaptive wearable ankle robot for the treatments of ankle sprain through physical rehabilitation. The ankle robot has a bioinspired design, devised after a careful study of the improvement opportunities in the existing ankle robots. Robot design is adaptable to subjects of varying physiological abilities and age groups. Ankle robot employs lightweight but powerful pneumatic muscle actuators (PMA) which mimics skeletal muscles in actuation. To address nonlinear characteristics of PMA, a fuzzy-based disturbance observer (FBDO) has been developed. Another instance of an adaptive fuzzy logic controller based on Mamdani inference has been developed and appended with the FBDO to compensate for the transient nature of the PMA. With the proposed control scheme, it is possible to simultaneously control four parallel actuators of the ankle robot and achieve three rotational degrees of freedom. To evaluate the robot design, the disturbance observer, and the adaptive fuzzy logic controller, experiments were performed. The ankle robot was used by a neurologically intact subject. The robot-human interaction was kept as active-passive while the robot was operated on predefined trajectories commonly adopted by the therapists. Trajectory tracking results are reported in the presence of an unpredicted human user intervention, use of compliant and nonlinear actuators, and parallel kinematic structure of the ankle robot.
Keywords :
adaptive control; compliant mechanisms; design engineering; fuzzy control; fuzzy reasoning; human-robot interaction; injuries; medical robotics; muscle; neurophysiology; observers; patient rehabilitation; patient treatment; pneumatic actuators; trajectory control; FBDO; Mamdani inference; PMA nonlinear characteristics; adaptive fuzzy logic controller; adaptive wearable parallel ankle robot; age groups; ankle injury treatment; ankle sprain treatment; bioinspired design; compliant actuators; fuzzy-based disturbance observer; human user intervention; neurologically intact subject; nonlinear actuators; parallel actuators; parallel kinematic structure; physical rehabilitation; physiological abilities; pneumatic muscle actuators; robot design; robot-human interaction; rotational degrees of freedom; skeletal muscles; trajectory tracking; Actuators; Injuries; Joints; Muscles; Robot kinematics; Robot sensing systems; Adaptive fuzzy logic controller; fuzzy-based disturbance observer (FBDO); pneumatic muscle actuators (PMAs); wearable ankle rehabilitation robot;
fLanguage :
English
Journal_Title :
Mechatronics, IEEE/ASME Transactions on
Publisher :
ieee
ISSN :
1083-4435
Type :
jour
DOI :
10.1109/TMECH.2012.2219065
Filename :
6327674