• DocumentCode
    544275
  • Title

    A controllable wide band hyperthermia applicator

  • Author

    Rajhi, Adnen ; Takagi, Tasuku

  • Author_Institution
    Dept. Electr. Commu., Tohoku Univ., Sendai, Japan
  • Volume
    1
  • fYear
    1992
  • fDate
    Oct. 29 1992-Nov. 1 1992
  • Firstpage
    249
  • Lastpage
    251
  • Abstract
    This paper treats a new kind of hyperthermia applicator which is designed to combine the powers from two different EM source arrays coupled through a cylindrical cavity; *one source array has the ability to control the operating frequency by adjusting the partially filled cylindrical cavity length (air gap). *the other source array (aperture array) is designed under the operating condition of the previous source. The experiment structure employs coaxial feeding and aperture feedings through waveguides. According to the radius of the cylindrical cavity up to four coaxial feeding and four waveguides has been combined to operate at 2450Mhz.
  • Keywords
    biomedical equipment; hyperthermia; microwave devices; waveguides; EM source arrays; air gap; aperture array; aperture feeding; coaxial feeding; controllable wide band hyperthermia applicator; frequency 2450 MHz; operating frequency control; partially filled cylindrical cavity length; waveguides; Biology; Cavity resonators; Thickness measurement;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 1992 14th Annual International Conference of the IEEE
  • Conference_Location
    Paris
  • Print_ISBN
    0-7803-0785-2
  • Electronic_ISBN
    0-7803-0816-6
  • Type

    conf

  • DOI
    10.1109/IEMBS.1992.5760949
  • Filename