  • Title

    SAR distributions of a dielectric waveguide applicator

  • Author

    Tanaka, Ryoji ; Nikawa, Yoshio ; Mori, Shinsaku

  • Author_Institution
    Dept. of Electr. Eng., Keio Univ., Yokohama, Japan
  • Volume
    1
  • fYear
    1992
  • fDate
    Oct. 29 1992-Nov. 1 1992
  • Firstpage
    252
  • Lastpage
    253
  • Abstract
    Applying dielectric waveguide for applicator in hyperthermia treatment, an applicator having no metal walls can be made, which has possibility to be small and flexible. In the practical treatment, it is very important to be previously informed the distribution of specific absorption rate (SAR) in order to perform hyperthermia treatment in safety. The finite-different time-domain (FD-TD) technique is used to calculate SAR distribution in the human body using the experimental data of EM field generated by a dielectric waveguide applicator.
  • Keywords
    biomedical equipment; dosimetry; finite difference time-domain analysis; hyperthermia; waveguides; FD-TD technique; SAR distribution calculation; dielectric waveguide applicator SAR distribution; dielectric waveguide applicator generated EM field; finite difference time domain technique; human body; hyperthermia treatment; specific absorption rate; Applicators; Nonhomogeneous media;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 1992 14th Annual International Conference of the IEEE
  • Conference_Location
    Paris
  • Print_ISBN
    0-7803-0785-2
  • Electronic_ISBN
    0-7803-0816-6
  • Type

    conf

  • DOI
    10.1109/IEMBS.1992.5760950
  • Filename
    5760950