DocumentCode :
544276
Title :
SAR distributions of a dielectric waveguide applicator
Author :
Tanaka, Ryoji ; Nikawa, Yoshio ; Mori, Shinsaku
Author_Institution :
Dept. of Electr. Eng., Keio Univ., Yokohama, Japan
Volume :
1
fYear :
1992
fDate :
Oct. 29 1992-Nov. 1 1992
Firstpage :
252
Lastpage :
253
Abstract :
Applying dielectric waveguide for applicator in hyperthermia treatment, an applicator having no metal walls can be made, which has possibility to be small and flexible. In the practical treatment, it is very important to be previously informed the distribution of specific absorption rate (SAR) in order to perform hyperthermia treatment in safety. The finite-different time-domain (FD-TD) technique is used to calculate SAR distribution in the human body using the experimental data of EM field generated by a dielectric waveguide applicator.
Keywords :
biomedical equipment; dosimetry; finite difference time-domain analysis; hyperthermia; waveguides; FD-TD technique; SAR distribution calculation; dielectric waveguide applicator SAR distribution; dielectric waveguide applicator generated EM field; finite difference time domain technique; human body; hyperthermia treatment; specific absorption rate; Applicators; Nonhomogeneous media;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Engineering in Medicine and Biology Society, 1992 14th Annual International Conference of the IEEE
Conference_Location :
Paris
Print_ISBN :
0-7803-0785-2
Electronic_ISBN :
0-7803-0816-6
Type :
conf
DOI :
10.1109/IEMBS.1992.5760950
Filename :
5760950
Link To Document :
بازگشت