Abstract :
A Monte Carlo simulation has been developed to predict the quality of time resolved images of the breast by transillumination. The smallest diameter of a detectable carcinoma located in the breast has been computed. The simulation suggests that time resolved imaging of the breast is possible and invaluable in the near infra-red (NIR) by transillumination. The estimated diameter of the smallest detectable sphere is de rived from the image quality index theory (or IQI theory) and its value is around 4 mm.
Keywords :
Monte Carlo methods; biological organs; biomedical optical imaging; high-speed optical techniques; infrared imaging; mammography; IQI theory; Monte Carlo simulation; breast; carcinoma; image quality index theory; time resolved imaging; time resolved transillumination; Analytical models; Biomedical imaging; Image resolution; Indexes; X-rays;