• DocumentCode
    544454
  • Title

    Transoesophageal defibrillation

  • Author

    Anderson, J.A. ; Croal, S ; Mckeown, P ; Cochrane, D J ; Adgey, A.A.J.

  • Author_Institution
    N. Ireland Bio-Engineering Centre, University of Ulster at Jordanstown, Shore Road, Co. Antrim BT37 OQB, N. Ireland
  • Volume
    2
  • fYear
    1992
  • fDate
    Oct. 29 1992-Nov. 1 1992
  • Firstpage
    667
  • Lastpage
    668
  • Abstract
    The conventional approach to external trans-chest deflbrillation is an extremely low efficient method of delivering energy to the heart. The oesophagus shares a close anatomical relationship with the heart and offers a more optimal pathway for the delivery of a DC countershock. This paper reports on the design of a new unipolar catheter for transoesophageal deflbrillation. The work includes animal data followed by a series of human studies.
  • Keywords
    Area measurement; Cardiology; Defibrillation; Impedance; Impedance measurement;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 1992 14th Annual International Conference of the IEEE
  • Conference_Location
    Paris, France
  • Print_ISBN
    0-7803-0785-2
  • Electronic_ISBN
    0-7803-0816-6
  • Type

    conf

  • DOI
    10.1109/IEMBS.1992.5761162
  • Filename
    5761162