DocumentCode
544492
Title
Simulation study of optimal configuration and control of multi-chamber balloon for IABP
Author
Bai, Jing ; Lin, Hang
Author_Institution
Department of Electrical Engineering, Tsinghua University, Beijing, P.R. China
Volume
2
fYear
1992
fDate
Oct. 29 1992-Nov. 1 1992
Firstpage
752
Lastpage
753
Abstract
Balloon configuration and control scheie are iiportant for the optimization of the assist to the failing heart Kith IABP. In this Hork, the configuration of lulti-chaiber balloons and their control scheies have been investigated by using a heiodynaiic lodel and computer simulation lethods. Results show that the optiial configuration and controlled lulti-chaibered balloon can provide better assist to the failing heart. Based on the coiputer simulation, it Mas found that the shape of the rear chaiber in lulti-chaibered balloon is critical. The optiial control scheie for this rear chaiber should be to inflate it first and deflate it last.
Keywords
Blood; Cardiology; Heart rate; Hemodynamics; Simulation; Thyristors;