    Selective recording of thin muscular layers with improved Ag-AgCI electrodes

  • Author

    Crenner, F. ; Angel, F. ; Ringwald, C.

  • Author_Institution
    INSERM Unit 61, Surgery B, Civil Hospital F-67000 Strasbourg- France
  • Volume
    6
  • fYear
    1992
  • fDate
    Oct. 29 1992-Nov. 1 1992
  • Firstpage
    2366
  • Lastpage
    2367
  • Abstract
    An electrode assembly has been developed to record in vivo in the animal, simultaneously from both sides of the device, the electromyogram (EMG) of different muscular layers of the intestinal wall. The transducer is composed of two opposite active electrodes, and is surrounded by a ring electrode which is used as reference and avoids recording of interfering signals. An improved chlorination technique of silver electrodes, using a partial electrolytic removal of the chloride, is presented : it decreases the electrodes impedance and lowers their ageing degradation. Techniques of evaluating the performance of electrodes by measurement of their complex impedance are critically discussed.
  • Keywords
    Current measurement; Electrodes; Frequency measurement; Impedance measurement; Phase measurement; Transducers; Wires;