• DocumentCode
    544669
  • Title

    Microprocessor controlled bioimpedance measurements

  • Author

    Schmukler, Robert E. ; Dixon, Jonathan W.

  • Author_Institution
    Electrophysics Branch, Center for Devices and Radiological Health, OST, FDA 12721 Twinbrook Parkway, Rockville, MD 20857
  • Volume
    6
  • fYear
    1992
  • fDate
    Oct. 29 1992-Nov. 1 1992
  • Firstpage
    2381
  • Lastpage
    2381
  • Abstract
    This paper describes the measurements made possible by using microprocessor-control to operate a Solartron 1260 Impedance Analyzer. Calibration experiments were able to demonstrate that an apparent two time-constant impedance measurement is really only a one time-constant system. This is due to artifacts introduced by the parasitic capacitance and the source resistance to the reference plane of the measurement system.
  • Keywords
    Impedance measurement; Measurement uncertainty;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 1992 14th Annual International Conference of the IEEE
  • Conference_Location
    Paris, France
  • Print_ISBN
    0-7803-0785-2
  • Electronic_ISBN
    0-7803-0816-6
  • Type

    conf

  • DOI
    10.1109/IEMBS.1992.5761413
  • Filename