DocumentCode
544669
Title
Microprocessor controlled bioimpedance measurements
Author
Schmukler, Robert E. ; Dixon, Jonathan W.
Author_Institution
Electrophysics Branch, Center for Devices and Radiological Health, OST, FDA 12721 Twinbrook Parkway, Rockville, MD 20857
Volume
6
fYear
1992
fDate
Oct. 29 1992-Nov. 1 1992
Firstpage
2381
Lastpage
2381
Abstract
This paper describes the measurements made possible by using microprocessor-control to operate a Solartron 1260 Impedance Analyzer. Calibration experiments were able to demonstrate that an apparent two time-constant impedance measurement is really only a one time-constant system. This is due to artifacts introduced by the parasitic capacitance and the source resistance to the reference plane of the measurement system.