DocumentCode
544670
Title
A method for the measurement of the properties of individual electrode-skin interfaces and the implications of the electrode properties for preamplifier design
Author
Grimbergen, C.A. ; VanRijn, A.C.Metting ; Peper, A.
Author_Institution
Medical Physics & Informatics Dept., University of Amsterdam, Meibergdreef 15, 1105 AZ Amsterdam, The Netherlands
Volume
6
fYear
1992
fDate
Oct. 29 1992-Nov. 1 1992
Firstpage
2382
Lastpage
2383
Abstract
A method to measure impedances of individual electrodes has been developed. The DC- and AC-loading conditions of the electrode can be varied in a large range. Impedance values of electrodes on unprepared or alcohol rubbed skin are measured to be very high and vary between electrodes both on the same test person and on different test persons. This leads to important consequences for interference reduction and current noise properties of the bioelectric amplifier.
Keywords
Capacitance; Impedance; Laboratories; Voltage measurement;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, 1992 14th Annual International Conference of the IEEE
Conference_Location
Paris, France
Print_ISBN
0-7803-0785-2
Electronic_ISBN
0-7803-0816-6
Type
conf
DOI
10.1109/IEMBS.1992.5761414