Mechanical properties of catheters: Designing of tests for kinking, traction, and torsion and their validation on a series of central venous catheters
Author :
Bailly, A.-L. ; Debout, J. ; Laccourreye, O. ; Laurent, A. ; Merland, J.J.
Author_Institution :
Dpt of Interventional Neuroradiology, Claude Bernard Res. Fund, Paris, France
Volume :
3
fYear :
1992
fDate :
Oct. 29 1992-Nov. 1 1992
Firstpage :
1162
Lastpage :
1163
Abstract :
This paper describes three tests that have been designed and adapted in our laboratory to test the mechanical properties of catheter tubes in their usable form, to predict their mechanical behaviour In vivo. They have been validated on a series of 86 central venous catheters and their dlscrimitant power and interecorreIations compared.
Keywords :
biomechanics; catheters; traction; central venous catheters; dlscrimitant power; kinking; mechanical properties; torsion; traction; Catheters; In vivo; Laboratories; Mechanical factors; Mechanical variables measurement; Polymers;