• DocumentCode
    544746
  • Title

    A piezoelectric belt for cardiac pulse and respiration measurements on small mammals

  • Author

    Aminian, K. ; Thouvenin, X. ; Robert, Ph. ; Seydoux, J. ; Girardier, L.

  • Author_Institution
    Laboratoire de Métrologie, Federal Institute of Technology, 1015 Lausanne - Switzerland
  • Volume
    6
  • fYear
    1992
  • fDate
    Oct. 29 1992-Nov. 1 1992
  • Firstpage
    2663
  • Lastpage
    2664
  • Abstract
    A non Invasive sensor for measuring both the respiration and cardiac rates on small mammals is described. This sensor, designed as a belt, involves four piezoelectric films (PVDF) devoted to measuring the pressure changes at several points of the thoracic cage. This paper focuses on mechanical design as well as theoretical and practical responses of the sensors developed. An experimental result, carried out on an anaesthetised rat is also presented.
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 1992 14th Annual International Conference of the IEEE
  • Conference_Location
    Paris, France
  • Print_ISBN
    0-7803-0785-2
  • Electronic_ISBN
    0-7803-0816-6
  • Type

    conf

  • DOI
    10.1109/IEMBS.1992.5761499
  • Filename
    5761499