DocumentCode
544807
Title
First approach for respiratory monitoring by amplitude demodulation of the electrocardiogram
Author
Khaled, Ziad Bou ; Farges, Gilbert
Author_Institution
URA 858, Université de Technologie de Compiegne, B.P. 649,60206 Compiègne, Cedex, France
Volume
6
fYear
1992
fDate
Oct. 29 1992-Nov. 1 1992
Firstpage
2535
Lastpage
2536
Abstract
This paper presents a new method for cardiorespiratory monitoring called Cardio-Modulo-Respirography (C.M.R.). Its principle is based on amplitude demodulation of the electrocardiogram (ECG). The C.M.R. is compared to the classical method of Impedance Rheography (I.R.) by simultaneous respiratory signal recordings on newborns and adults. On a qualitative analysis, C.M.R. is more insensitive to movement- and cardiac artifacts than I.R. Furthermore, it permits the cardiorespiratory monitoring and apnea detection without electric current injection throuth the chest. The first results lead us to improve the C.M.R. electronic system sensibility in order to adjust it to new-borns and prematures.
Keywords
Cardiology;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, 1992 14th Annual International Conference of the IEEE
Conference_Location
Paris, France
Print_ISBN
0-7803-0785-2
Electronic_ISBN
0-7803-0816-6
Type
conf
DOI
10.1109/IEMBS.1992.5761575
Filename
5761575
Link To Document