Estimating volumes of intra-abdominal blood using electrical impedance imaging
Author :
Sadleir, R.J. ; Fox, R.A. ; van Kann, F.J. ; Attikiouzel, Y.
Author_Institution :
Departments of Physics and Electrical Engineering, University of Western Australia, Nedlands 6009, Western Australia
Volume :
5
fYear :
1992
fDate :
Oct. 29 1992-Nov. 1 1992
Firstpage :
1750
Lastpage :
1751
Abstract :
Electrical Impedance Imaging is a non-invasive method of imaging which we intend to apply to the problem of detecting intraabdominal bleeding in Emergency Centre patients. Experiments performed using a cylindrical saline-gel phantom show that as little as 25 ml of free blood within the abdomen may be detected. We also find a strong linear correlation between the total resistivity change created by the anomaly and the volume of the equivalent blood anomaly; thus enabling us to easily quantify the volume of blood perfused into the abdomen over a period of time directly from a reconstructed image. The total resistivity change produced by an anomaly is also observed to be independent of its radial position within the phantom.
Keywords :
Humans; Image reconstruction; Impedance; Phantoms;
fLanguage :
English