Title :
Design of a multicrystal detector array for PET using Monte Carlo techniques
Author :
Pavlopoulos, Sotiris ; Tzanakos, George
Author_Institution :
Department of Biomedical Engineering, Rutgers University, Piscataway, NJ 08855
fDate :
Oct. 29 1992-Nov. 1 1992
Abstract :
We have used Monte Carlo techniques to design a two-dimensional detector array for PET. The detector array consists of an 8×4 matrix of BGO crystals coupled to 4 photomultipliers. The design included optimizing the optical cross-talk between detectors in terms of positioning accuracy and detection efficiency. Some preliminary results including the energy resolution, the positioning accuracy and the detection efficiency of the module are presented.
Keywords :
Optical arrays; Optical crosstalk; Optical scattering; Positron emission tomography;
Conference_Titel :
Engineering in Medicine and Biology Society, 1992 14th Annual International Conference of the IEEE
Conference_Location :
Paris, France
Print_ISBN :
0-7803-0785-2
Electronic_ISBN :
0-7803-0816-6
DOI :
10.1109/IEMBS.1992.5762232