Title :
Evaluating periodicity of trabecular bone phantoms using ultrasound signals
Author :
Rosa, P.T.C.R. ; Matusin, D.P. ; Pereira, A.J.F. ; Machado, C.B. ; Pereira, W.C.A.
Author_Institution :
Biomed. Eng. Program, Fed. Univ. of Rio de Janeiro, Rio de Janeiro, Brazil
fDate :
March 28 2011-April 1 2011
Abstract :
The aim of this work was to characterize the periodicity of a commercial available trabecular bone phantom with ultrasound (US) backscattered signals. Three US frequencies were applied (1.0, 2.25 and 5.0 MHz), and spectral analysis methods were used to estimate the mean scatterer spacing (MSS) of 100 signals/sample from five different samples. The results showed that it is possible to estimate the medium periodicity (trabecular spacings varying from 1.5 to 2.5 mm), and at higher frequencies the average MSS decreased, which is expected because of a higher spatial resolution (identification of the trabecular thickness periodicity). MSS values are in agreement with intertrabecular distances declared by the product supplier.
Keywords :
backscatter; biomedical ultrasonics; bone; phantoms; ultrasonic scattering; intertrabecular distances; mean scatterer spacing; periodicity; spectral analysis methods; trabecular bone phantoms; ultrasound backscattered signals; ultrasound signals; Acoustics; Bones; Conferences; Frequency estimation; IEEE catalog; Phantoms; Ultrasonic imaging; Mean scatterer spacing; periodicity; phantom; ultrasound;
Conference_Titel :
Health Care Exchanges (PAHCE), 2011 Pan American
Conference_Location :
Rio de Janeiro
Print_ISBN :
978-1-61284-915-7
DOI :
10.1109/PAHCE.2011.5871911