  • Abstract
    The aim of this work was to characterize the periodicity of a commercial available trabecular bone phantom with ultrasound (US) backscattered signals. Three US frequencies were applied (1.0, 2.25 and 5.0 MHz), and spectral analysis methods were used to estimate the mean scatterer spacing (MSS) of 100 signals/sample from five different samples. The results showed that it is possible to estimate the medium periodicity (trabecular spacings varying from 1.5 to 2.5 mm), and at higher frequencies the average MSS decreased, which is expected because of a higher spatial resolution (identification of the trabecular thickness periodicity). MSS values are in agreement with intertrabecular distances declared by the product supplier.
  • Keywords
    backscatter; biomedical ultrasonics; bone; phantoms; ultrasonic scattering; intertrabecular distances; mean scatterer spacing; periodicity; spectral analysis methods; trabecular bone phantoms; ultrasound backscattered signals; ultrasound signals; Acoustics; Bones; Conferences; Frequency estimation; IEEE catalog; Phantoms; Ultrasonic imaging; Mean scatterer spacing; periodicity; phantom; ultrasound;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Health Care Exchanges (PAHCE), 2011 Pan American
  • Conference_Location
    Rio de Janeiro
  • Print_ISBN
    978-1-61284-915-7
  • Type

    conf

  • DOI
    10.1109/PAHCE.2011.5871911
  • Filename
    5871911