• DocumentCode
    55257
  • Title

    Direct Comparison of Josephson Voltage Standards at 10 V Between BIPM and CENAM

  • Author

    Aviles, David ; Navarrete, Enrique ; Hernandez, D. ; Solve, S. ; Chayramy, R.

  • Author_Institution
    Centro Nac. de Metrol., Querétaro, France
  • Volume
    62
  • Issue
    6
  • fYear
    2013
  • fDate
    Jun-13
  • Firstpage
    1640
  • Lastpage
    1645
  • Abstract
    A direct comparison of Josephson voltage standards at 10 V between the Bureau International des Poids et Mesures (BIPM) and the Centro Nacional de Metrología (CENAM), México, was carried out in September 2011. This comparison is part of the BIPM key comparisons (BIPM.EM-K10.b) and took place in the new DC Voltage Laboratory of CENAM. Both systems were very stable during the comparison, and the type “A” uncertainty was very low. The relative voltage difference between the two quantum standards was -6 parts in 1011 ( -0.6 nV), with a combined uncertainty of 7 parts in 1011 (0.7 nV).
  • Keywords
    measurement standards; measurement uncertainty; voltage measurement; BIPM; Bureau International des Poids et Mesures; CENAM; Centro Nacional de Metrología; DC Voltage Laboratory; Josephson voltage standard; México; measurement uncertainty; quantum standard; relative voltage difference; voltage -0.6 nV; voltage 0.7 nV; voltage 10 V; Detectors; Frequency measurement; Microwave measurements; Radiation detectors; Standards; Uncertainty; Voltage measurement; DC voltage measurement; DC voltage standard measurements; JVS comparison; Josephson voltage standard (JVS); standards comparison;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/TIM.2012.2228752
  • Filename