• DocumentCode
    55369
  • Title

    First K-Edge Imaging With a Micro-CT Based on the XPAD3 Hybrid Pixel Detector

  • Author

    Cassol Brunner, F. ; Dupont, Matthieu ; Meessen, C. ; Boursier, Y. ; Ouamara, Hamid ; Bonissent, A. ; Kronland-Martinet, C. ; Clemens, Joachim ; Debarbieux, Franck ; Morel, Carine

  • Author_Institution
    CPPM, Aix-Marseille Univ., Marseille, France
  • Volume
    60
  • Issue
    1
  • fYear
    2013
  • fDate
    Feb. 2013
  • Firstpage
    103
  • Lastpage
    108
  • Abstract
    We investigate the capability to perform K-edge imaging with the newly developed micro-CT PIXSCAN based on the XPAD3 hybrid pixel detector. The XPAD3 detector surface of 8 cm ×11 cm makes it possible to perform whole body mouse imaging. We present a proof of principle of K-edge imaging of mouse-size phantoms filled with Silver and Iodine solutions. Results are compared with standard X-ray absorption tomography for several solution densities.
  • Keywords
    X-ray detection; X-ray imaging; computerised tomography; phantoms; semiconductor counters; K-edge imaging; XPAD3 detector surface; XPAD3 hybrid pixel detector; iodine solutions; microCT PIXSCAN based detector; mouse-size phantoms; silver solutions; standard X-ray absorption tomography; Detectors; Electron tubes; Materials; Phantoms; Photonics; Silver; Hybrid integrated circuits; X-ray tomography; image enhancement; pixel sensors;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.2012.2217753
  • Filename
    6329459