Title :
A simulation study of dynamic MRI based on partially separable functions and keyhole techniques
Author :
Xie, Guo-Xi ; Weng, Zhuo ; Zou, Chao ; Feng, Xiang ; Zheng, Hai-Rong ; Liu, Xin ; Qiu, Bensheng
Author_Institution :
Inst. of Biomed. & Health Eng., Paul C. Lauterbur Res. Center for Biomed. Imaging, Shenzhen, China
Abstract :
The partially separable function (PSF) model can significantly improve MRI spatial and temporal resolutions. However, the pre-data acquisition of PSF is very time-consuming in practical. To overcome this limitation, in this paper we proposed using keyhole for fast pre-data acquisition of PSF. The simulation results showed that the pre-data acquired by keyhole can be properly processed by partially separable functions, and the MR images can be reconstructed in high quality without any motion artifacts. This result demonstrated the potential of dynamic MRI by combine use of PSF and keyhole techniques.
Keywords :
biomedical MRI; image reconstruction; image resolution; medical image processing; dynamic MRI; image reconstruction; keyhole techniques; motion artifacts; partially separable functions; predata acquisition; spatial resolution; temporal resolution; Image reconstruction; Magnetic resonance imaging; Navigation; Signal to noise ratio; Spatial resolution; MRI; keyhole; partially separable functions;
Conference_Titel :
Biomedical Engineering and Informatics (BMEI), 2011 4th International Conference on
Conference_Location :
Shanghai
Print_ISBN :
978-1-4244-9351-7
DOI :
10.1109/BMEI.2011.6098242