  • Title

    A fluoroscopy-based navigation system for fixation of femoral neck fracture

  • Author

    Yang, Yan ; Hu, Lei ; Wang, Junchen

  • Author_Institution
    Robotic Inst., Beihang Univ., Beijing, China
  • Volume
    1
  • fYear
    2011
  • fDate
    15-17 Oct. 2011
  • Firstpage
    24
  • Lastpage
    28
  • Abstract
    In this paper, a new fluoroscopy-based navigation system was developed. It can track the surgery instrument and patient´s femur´s location real-timely. The preoperative planning was against three cannulated screws based on the principle of close reduction and internal fixation (CRIF) treatment. To project the surgical instrument into the two fluoroscopic images (AP and Literal) of the femoral neck, firstly we calibrated the C-arm camera and then made a hand-eye calibration against the C-arm system. Finally, using the transformation matrix of different coordinates realized the navigation process. The experiment shows that the accuracy of system was blow 2mm. The navigation system was proved to be functional and less radiation exposure.
  • Keywords
    biomedical equipment; bone; calibration; cameras; diagnostic radiography; surgery; C-arm camera; close reduction internal fixation treatment; femoral neck fracture fixation; femur; fluoroscopy based navigation system; hand-eye calibration; surgery instrument; transformation matrix; Calibration; Cameras; Fasteners; Instruments; Navigation; Neck; Surgery; Computer assisted surgery; cannulated screws; femoral neck fracture; fluoroscopy-based navigation system;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Biomedical Engineering and Informatics (BMEI), 2011 4th International Conference on
  • Conference_Location
    Shanghai
  • Print_ISBN
    978-1-4244-9351-7
  • Type

    conf

  • DOI
    10.1109/BMEI.2011.6098336
  • Filename