  • Abstract
    Knowledge of the biomechanical behavior of the knee prosthesis can have an important influence on balance, stabilization and lifetime of total knee prosthesis. A ligament tensioning device with pressure sensor was used to measure the torque versus rotation between the tibia and femur during total knee arthroplasties. The measurements were performed with the knee flexed, after balance in extension. We followed the testing of a mathematical model description, evaluation of the force / torque accumulated in the capsular ligament structures during balancing and at 5 ° of knee deviation. Statistical tests have validated the model. The force balance was influenced by the inclination of the tibial surface, the extension deficit, the size of the tibial plateau, the deviation of HKA axis and slice thickness of tibial osteotomy. These data help to establish the force necessary for joint replacement. They demonstrate that the force required during prosthesis balancing is determined by many factors and thus it´s incalculable preoperatively. This requires the use of devices that indicate equilibration intraoperatively.
  • Keywords
    biomechanics; bone; mathematical analysis; mechanoception; pressure sensors; statistical analysis; tensors; torque; biomechanical behavior; capsular ligament structures; femur; force balance; intraoperative joint capsule; joint replacement; knee prosthesis; ligament tensioning device; mathematical model description; posterior stabilized TKA; pressure sensor tensor; statistical tests; tibia; tibial osteotomy; Biomechanics; Force; Knee; Ligaments; Mathematical model; Surgery; Torque; distraction force and posterior stabilized total knee arthroplasty; knee biomechanics; ligament balance; mathematical model; pressure sensor;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    E-Health and Bioengineering Conference (EHB), 2011
  • Conference_Location
    Iasi
  • Print_ISBN
    978-1-4577-0292-1
  • Type

    conf

  • Filename
    6150348