• DocumentCode
    561469
  • Title

    Doppler radar for heartbeat rate and heart rate variability extraction

  • Author

    Obeid, D. ; Sadek, S. ; Zaharia, G. ; Zein, G. El

  • Author_Institution
    Kaptalia Monitoring, Vannes, France
  • fYear
    2011
  • fDate
    24-26 Nov. 2011
  • Firstpage
    1
  • Lastpage
    4
  • Abstract
    This paper presents a Doppler radar system used to detect the heartbeat signal from a distance of one meter. The proposed system is based on using a vector network analyzer and two antennas. Measurements are performed at 16 GHz for different power levels between 0 and -25 dBm. Both heartbeat rate and heart rate variability are extracted and compared to a simultaneous ECG signal.
  • Keywords
    Doppler radar; biomedical telemetry; cardiology; network analysers; radar antennas; Doppler antennas; Doppler radar; ECG signal comparison; frequency 16 GHz; heart rate variability extraction; heartbeat rate extraction; vector network analyzer; Doppler radar; Electrocardiography; Heart rate variability; Microwave communication; Smoothing methods; Doppler effect; cardiopulmonary activity; heartbeat rate; microwave systems; noncontact detection;