DocumentCode
564386
Title
Initial evaluation of the USNO rubidium fountain
Author
Peil, S. ; Crane, S. ; Swanson, T. ; Ekstrom, C.R.
Author_Institution
Clock Dev. Div., U.S. Naval Obs., Washington, DC, USA
fYear
2006
fDate
27-30 March 2006
Firstpage
194
Lastpage
196
Abstract
We discuss the first rubidium atomic fountain at the U. S. Naval Observatory, NRF1, which has been built as a prototype for future devices to be included in the USNO timescale. The system has demonstrated a short-term Allan deviation of 1.35×1013/τ1/2 when measured against a hydrogen maser. We have directly compared NRF1´s performance to that of our cesium fountain, NCF, demonstrating integration as 1/τ1/2 below 1×10-15, and performance that is consistent with 7×10-16 at 11 hours.
Keywords
atomic clocks; masers; rubidium; NRF1; Rb; U. S. Naval Observatory; USNO rubidium atomic fountain clock; cesium fountain clock; hydrogen maser; short-term Allan deviation; time 11 hour; Atomic clocks; Crystals; Frequency measurement; Hydrogen; Masers; Noise;
fLanguage
English
Publisher
ieee
Conference_Titel
Frequency and Time Forum (EFTF), 2006 20th European
Conference_Location
Braunschweig
Print_ISBN
978-1-4673-2642-1
Type
conf
Filename
6230970
Link To Document