    A low cost instrumentation system has been proposed in this paper to determine the stability of dental implants. For this purpose, Resonance Frequency Analysis technique has been used to monitor the changes in resonance frequency that takes place during the bonding of the dental implant with the surrounding bones and tissues [1]. In the present work, the designed instrumentation system has been characterized and its performance validated in in-vitro experiments. From this study, it is observed that the instrumentation system developed can be used for determining stability of dental implants. There is, however, scope for improvements in terms of the resolution and the real-time performance of the system.
  • Keywords
    biomedical equipment; bone; dentistry; prosthetics; biological tissues; bonding; bones; dental implant stability; instrumentation system; resonance frequency analysis; Circuit stability; Dentistry; Impedance; Implants; Instruments; Resonant frequency; Stability analysis; Resonance frequency analysis; dental implant stability; impedance response; instrumentation system; phase detector circuit;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Emerging Applications of Information Technology (EAIT), 2012 Third International Conference on
  • Conference_Location
    Kolkata
  • Print_ISBN
    978-1-4673-1828-0
  • Type

    conf

  • DOI
    10.1109/EAIT.2012.6407867
  • Filename
    6407867