  • Abstract
    The purpose of this paper is to assess electromagnetic environment of nuclear magnetic resonance (NMR) device in hospital, in considering of the proposed construction of subway lines which would pass the hospital. A method is proposed to quantitatively analyze and evaluate electromagnetic interference of NMR device. It is derived through the analysis and evaluation: when subway line passes through the hospital, the field strength surrounding the hospital can reach 30dBuT to 40dBuT at several frequencies, which is greater than 20dBuT-the threshold of the NMR device. It can ensure the NMR equipment working normally, after applying an electromagnetic shielding measure which makes sure that the electromagnetic shielding effectiveness reaches 80dB.
  • Keywords
    biomedical NMR; biomedical equipment; electromagnetic interference; electromagnetic shielding; hospitals; NMR device; NMR equipment; electromagnetic environment assessment; electromagnetic interference; electromagnetic shielding measure; field strength; hospital; nuclear magnetic resonance device; subway lines; Analytical models; Antennas; Green products; Hospitals; Magnetic fields; Nuclear magnetic resonance; NMR device; electromagnetic environment; test;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Antennas, Propagation & EM Theory (ISAPE), 2012 10th International Symposium on
  • Conference_Location
    Xian
  • Print_ISBN
    978-1-4673-1799-3
  • Type

    conf

  • DOI
    10.1109/ISAPE.2012.6408967
  • Filename
    6408967