Abstract :
This paper presents a miniaturized brain cortex cooling device and system using thermoelectric (TE) cooler for hypothermia therapy application on freely moving rat. Two type of cooling electrodes including planar and micro-needle structures are fabricated and studied by EEG recording. Comparing to prior arts, our approach can be applied under freely moving behavior. The cooling temperature is designed as 2°C to avoid permanent injury in brain. Packaged planar or needle electrode with TE cooler consumes only 2×2mm2. EEG recording electrodes and thermocouple are implanted for observation as well. Simulation result shows that TE-cooler with needle type electrode behaves deeper effective cooling depth and faster cooling response in rodent cortex range (~2mm deep). Long term EEG monitoring on freely moving rat also presents no extra side effects with continuously cooling.
Keywords :
cooling; electrodes; patient treatment; EEG recording; cooling electrodes; freely moving rat; hypothermia therapy application; microneedle structure; miniaturized brain cortex cooling device; miniaturized cortex cooling device; planar structure; Brain modeling; Cooling; Electrodes; Epilepsy; Needles; Temperature measurement; Temperature sensors;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Sensors, 2012 IEEE
Conference_Location :
Taipei
ISSN :
1930-0395
Print_ISBN :
978-1-4577-1766-6
Electronic_ISBN :
1930-0395
Type :
conf
DOI :
10.1109/ICSENS.2012.6411273
Filename :
6411273
Link To Document :
بازگشت