Algorithm for systolic peak detection of pulse wave
Author
de la Vara Prieto, R.R.F. ; Rodriguez, D.D. ; Sanz, M.B.C. ; Mengana, A.S. ; Quintero, H.F.P.
Author_Institution
Centro de Biofisica Medica, Univ. de Oriente, Santiago de Cuba, Cuba
fYear
2012
fDate
1-5 Oct. 2012
Firstpage
1
Lastpage
4
Abstract
In this paper, a new method for systolic peak detection of pulse wave signals is presented. This method is based on the detection of each peak by the correlation with adaptive Gaussian function template (GFT). The point of maximum correlation is selected as a systolic peak. Its accuracy and noise robustness were evaluated over a set of annotated signals (arterial blood pressure) from CSL database. The GFT method showed an error respect to trained observers less than 5.5 ± 4.2 ms. Furthermore, this method has an error of 8.99 ± 25.09 ms for several noise realization at 9 dB of signal-noise ratio. The results suggest that this method could be used in the measurement of heart rate and pulse transit time, and in the study of autonomic function.
Keywords
Gaussian processes; medical signal detection; signal denoising; CSL database; GFT; adaptive Gaussian function template; arterial blood pressure; heart rate measurement; noise realization; pulse transit time measurement; pulse wave detection; pulse wave signal; signal-noise ratio; systolic peak detection; Accuracy; Band pass filters; Correlation; Databases; Noise robustness; Signal to noise ratio; Gaussian function; correlation; pulse wave; systolic peak;
fLanguage
English
Publisher
ieee
Conference_Titel
Informatica (CLEI), 2012 XXXVIII Conferencia Latinoamericana En
Conference_Location
Medellin
Print_ISBN
978-1-4673-0794-9
Type
conf
DOI
10.1109/CLEI.2012.6427234
Filename
6427234
Link To Document