• DocumentCode
    598845
  • Title

    Development of laser induced real time Photoacoustic Tomography imaging system and phantom evaluation

  • Author

    Shin, D. H. ; Ryu, Sang Hyun ; Song, C. G.

  • Author_Institution
    Div. of Electronics Eng., Chonbuk Natl. Univ., Korea
  • fYear
    2012
  • fDate
    21-24 Aug. 2012
  • Firstpage
    175
  • Lastpage
    179
  • Abstract
    Photoacoustic Tomography (PAT) is a promising medical imaging modality by reason of its particularity. It combines optical imaging contrast with the spatial resolution of ultrasound imaging, and can distinguish changes in biological features in an image. For these reasons, many studies are in progress to apply this technique for diagnosis. But, real-time PAT systems are necessary to confirm biological reactions induced by external stimulation immediately. Thus, we have developed a real-time PAT system using a linear array transducer and a custom-developed data acquisition board (DAQ). To evaluate the feasibility and performance of our proposed system, two types of phantom tests were also performed. As a result of those experiments, the proposed system shows satisfactory performance and its usefulness has been confirmed.
  • Keywords
    Linear array transducer; Photoacoustic Tomography; Real-time imaging;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Awareness Science and Technology (iCAST), 2012 4th International Conference on
  • Conference_Location
    Seoul, Korea (South)
  • Print_ISBN
    978-1-4673-2111-2
  • Electronic_ISBN
    978-1-4673-2110-5
  • Type

    conf

  • DOI
    10.1109/iCAwST.2012.6469609
  • Filename
    6469609