• DocumentCode
    605692
  • Title

    ECG signal acquisition within a continuous LF powered region

  • Author

    Franke, J. ; Heinen, U. ; Heidenreich, M. ; Morris, G.R. ; Buzug, T.

  • Author_Institution
    Bruker BioSpin MRI GmbH, Ettlingen, Germany
  • fYear
    2013
  • fDate
    23-24 March 2013
  • Firstpage
    1
  • Lastpage
    1
  • Abstract
    Neither unwanted MPI signals due to the presence of ECG electrode material within the sensitive Tx/Rx DF nor ECG signal perturbations caused by the 25 kHz DF excitation field have been detected for either electrode type. However, in case of electrode type [A] the maximum observed electrode temperature was 48.8 °C (c.f. Fig 1 LEFT) whereas a temperature of more than 115 °C (c.f. Fig. 1 RIGHT) has been detected in case of electrode type [B].
  • Keywords
    biomedical electrodes; electrocardiography; medical signal detection; ECG electrode material; ECG signal acquisition; ECG signal perturbations; MPI signals; continuous LF powered region; electrode temperature; frequency 25 kHz; temperature 48.8 degC; Animals; Electrocardiography; Electrodes; Magnetic resonance imaging; Monitoring; Temperature measurement;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Magnetic Particle Imaging (IWMPI), 2013 International Workshop on
  • Conference_Location
    Berkeley, CA
  • Print_ISBN
    978-1-4673-5520-9
  • Electronic_ISBN
    978-1-4673-5521-6
  • Type

    conf

  • DOI
    10.1109/IWMPI.2013.6528365
  • Filename
    6528365