DocumentCode :
61162
Title :
Individualized Estimation of the Central Aortic Blood Pressure Waveform: A Comparative Study
Author :
Jin-Oh Hahn
Author_Institution :
Dept. of Mech. Eng., Univ. of Maryland, College Park, MD, USA
Volume :
18
Issue :
1
fYear :
2014
fDate :
Jan. 2014
Firstpage :
215
Lastpage :
221
Abstract :
This paper presents a comparative study on the relative performance of alternative strategies for estimating an individualized central aortic blood pressure (BP) waveform. Based on the transmission line representation of the central aortic-radial arterial line, a fully individualized (ITF 1), two partially individualized (ITF 2 and ITF 3), and a fully nonindividualized (NITF) transfer functions (i.e., frequency-dependent central aortic-radial BP relationships) were constructed using experimental data collected from nine swine subjects. The central aortic BP waveforms estimated by these transfer functions were compared against their measured gold standards, with the root-mean-squared waveform error and the absolute errors associated with systolic and pulse pressures as performance measures. Overall, the advantage of the individualized over the nonindividualized approach was modest but significant. The superiority of the individualized approach to its nonindividualized counterpart was increasingly pronounced under nonnominal or extreme physiologic conditions, as the subject´s pulse transit time deviated from the averaged nominal value. The results suggest that the use of a fully individualized transfer function (ITF 1) is strongly recommended for nonnominal physiologic conditions, whereas a partially (ITF 2 and ITF 3) or even fully nonindividualized transfer function (NITF) may also yield acceptable performance under nominal physiologic conditions.
Keywords :
blood pressure measurement; blood vessels; cardiovascular system; electrocardiography; mean square error methods; medical signal processing; physiology; transfer functions; waveform analysis; absolute errors; central aortic blood pressure waveform; central aortic-radial arterial line; electrocardiography; extreme physiologic conditions; frequency-dependent central aortic-radial BP relationships; fully nonindividualized transfer functions; individualized estimation; nominal physiologic conditions; nonnominal physiologic conditions; pulse pressures; root-mean-squared waveform error; subject pulse transit time; systolic pressures; transmission line representation; Central aortic blood pressure; generalized transfer function; individualized transfer function;
fLanguage :
English
Journal_Title :
Biomedical and Health Informatics, IEEE Journal of
Publisher :
ieee
ISSN :
2168-2194
Type :
jour
DOI :
10.1109/JBHI.2013.2262945
Filename :
6516082
Link To Document :
بازگشت