• DocumentCode
    613492
  • Title

    Measurement of in-vivo patella kinematics using motion analysis and ultrasound (MAUS)

  • Author

    Monk, A.P. ; Chen, Mei ; Mellon, S. ; Gibbons, C.L.M.H. ; Beard, D.J. ; Gill, Harjot ; Murray, D.W.

  • Author_Institution
    NDORMS, Univ. of Oxford, Oxford, UK
  • fYear
    2013
  • fDate
    4-5 May 2013
  • Firstpage
    257
  • Lastpage
    260
  • Abstract
    This paper describes a motion analysis ultrasound system (MAUS) designed for the in-vivo study of the kinematics of patellofemoral joints (PFJ) in both normal and replaced knees. This system utilises non-iodising radiation to effectively acquire kinematic data during weight-bearing activities. Validation studies on a phantom established that the measurement accuracy of the system was 1.43 mm. A clinical validation trial is included.
  • Keywords
    biomedical ultrasonics; bone; image motion analysis; kinematics; medical image processing; phantoms; MAUS; PFJ; in-vivo patella kinematics; kinematic data acquisition; knee replacement; measurement accuracy; motion analysis ultrasound system; noniodising radiation; patellofemoral joints; phantom; weight-bearing activities; Calibration; Joints; Kinematics; Knee; Probes; Ultrasonic imaging; Ultrasonic variables measurement;