Title :
Use of an inductive divider for AC voltage ratio calibration
Author :
Draxler, Karel ; Styblikova, Renata
Author_Institution :
Dept. of Meas., CTU in Prague, Prague, Czech Republic
Abstract :
This paper presents the calibration of an electronic voltage divider and precise instrument voltage transformers in the range up to 1 kV at 50 Hz frequency using inductive voltage dividers. We also describe a way to evaluate the difference between a standard inductive divider and a calibrated instrument voltage transformer or electronic voltage divider expressed as the ratio error and phase displacement. The paper also presents the construction of a standard decadic voltage divider with a range of input voltage (10×100) V and (10×10) V.
Keywords :
calibration; potential transformers; voltage dividers; AC voltage ratio calibration; calibrated instrument voltage transformer; electronic voltage divider calibration; frequency 50 Hz; inductive voltage dividers; phase displacement; precise instrument voltage transformers; standard decadic voltage divider construction; Calibration; Magnetic circuits; Materials; Standards; Voltage measurement; Voltage transformers; Windings; inductive voltage divider; phase displacement; ratio error;
Conference_Titel :
Instrumentation and Measurement Technology Conference (I2MTC), 2013 IEEE International
Conference_Location :
Minneapolis, MN
Print_ISBN :
978-1-4673-4621-4
DOI :
10.1109/I2MTC.2013.6555398