  • Abstract
    New types of planar interdigital sensors have been fabricated on Silicon/Silicon Dioxide (Si/SiO2) wafers. The sensors were coated with pre-cursor silica functionalized with APTES (3-aminopropyltrietoxysilane) at different thicknesses. All sensors were then immobilized with Polymyxin, B (PmB). PmB is an antimicrobial peptide produced by the Gram-positive bacterium-Bacillus, has been immobilized on the coated sensors because of its specific binding properties to endotoxin. Studies were conducted to analyze the effect of different thicknesses of coatings on the sensitivity and selectivity of the sensors. It was observed sensors coated with 3 layers of coating has better sensitivity and selectivity to the target molecules (endotoxin) compared to sensors with 5 layers of coating. The repeatability and stability of the coated sensors were tested by multiple standard endotoxin measurement and it was observed that the sensors give a good reproducibility and stability up to six continuous measurements before the coating degrades.
  • Keywords
    chemical sensors; interdigital transducers; microorganisms; silicon; silicon compounds; 3-aminopropyltrietoxysilane; APTES; PmB; Si-SiO2; antimicrobial peptide; binding properties; coated sensor repeatability; coated sensor stability; coated sensors; coating thickness analysis; endotoxin detection; gram-positive bacterium-bacillus; planar interdigital sensors; polymyxin B; precursor silica; sensor selectivity; sensor sensitivity; silicon-silicon dioxide wafers; standard endotoxin measurement; Coatings; Current measurement; Electrodes; Impedance; Sensitivity; Sensors; Silicon; APTES; endotoxin; interdigital; polymyxin B; silicon dioxide;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Instrumentation and Measurement Technology Conference (I2MTC), 2013 IEEE International
  • Conference_Location
    Minneapolis, MN
  • ISSN
    1091-5281
  • Print_ISBN
    978-1-4673-4621-4
  • Type

    conf

  • DOI
    10.1109/I2MTC.2013.6555663
  • Filename
    6555663