• DocumentCode
    620356
  • Title

    Design of initiative lung simulator achieved by ARMls3ml138

  • Author

    Xiaomei Liu ; Linyuan Xu ; Lifeng Wei ; Shaochun Chen

  • Author_Institution
    Coll. of Inf. Eng., Shenyang Univ. of Chem. Technol., Shenyang, China
  • fYear
    2013
  • fDate
    25-27 May 2013
  • Firstpage
    3667
  • Lastpage
    3670
  • Abstract
    The problem of ventilators out of alignment or fault has been talked a lot in recent day. The basic solution has been described in this paper. To achieve comprehensive testing of ventilator an initiative lung simulator has been introduced here. A platform which is based on linear motor and ARMlm3s1138 processor has been build to solve this question. Some basic breath waveforms could be simulated by setting parameters on motion controller such as Respiratory rate, Ispiratory time, Tidal volume and Respiratory count. This simulator could be used to test and analyze whether a ventilator is reliable or not. The piston of anesthesia machine bellow gasbag is controlled to do reciprocating movement by linear motor, which is the basic principle of initiative lung simulator. After test, some basic breath state wave could be simulated by the lung simulator accurately.
  • Keywords
    biomedical equipment; linear motors; lung; medical control systems; microprocessor chips; motion control; pistons; ARMlm3s1138 processor; anesthesia machine bellow gasbag piston control; breath state wave simulation; breath waveform simulation; initiative lung simulator design; linear motor; motion controller; reciprocating movement; ventilators; Decision support systems; Linear motor and Respiratory waveform; Lung simulator; Ventilator;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Control and Decision Conference (CCDC), 2013 25th Chinese
  • Conference_Location
    Guiyang
  • Print_ISBN
    978-1-4673-5533-9
  • Type

    conf

  • DOI
    10.1109/CCDC.2013.6561585
  • Filename
    6561585