  • Abstract
    A 1 mm radius plane P(VDF-TrFE) transducer was designed and developed to calibrate hydrophone at high frequencies in far-field using a optical method. The transducer features a wide frequency range and a small effective radius. A membrane hydrophone was calibrated using the transducer in the frequency range from 10 MHz to 40 MHz in far-field with a 1 MHz frequency interval. The differences of the calibration results using the transducer and our current system were within 3 % in the frequency range from 10 MHz to 20 MHz.
  • Keywords
    biomedical transducers; biomedical ultrasonics; calibration; hydrophones; membranes; ultrasonic transducers; frequency 10 MHz to 40 MHz; membrane hydrophone calibration; optical method; plane P(VDF-TrFE) transducer; radius 1 mm; ultrasonic far-field; Acoustics; Calibration; Optical interferometry; Optical sensors; Sonar equipment; Transducers; Ultrasonic imaging;