DocumentCode :
620947
Title :
Imaging quality of the classical beamforming, SAFT and plane wave imaging - Experimental results
Author :
Karwat, P. ; Klimonda, Z. ; Lewandowski, Marcin ; Nowicki, Andrzej
Author_Institution :
Dept. of Ultrasound, Inst. of Fundamental Technol. Res. PAS, Warsaw, Poland
fYear :
2012
fDate :
7-10 Oct. 2012
Firstpage :
1
Lastpage :
4
Abstract :
The synthetic aperture focusing techniques (SAFT) are well known and widely deployed in radar techniques. Increasing processing power of modern computers allows effective implementation of various SAFT schemes in medical ultrasound systems with multi-element probes. The advance of the SAFT over the classical beamforming (BFR) is dynamic focusing in receive and transmit as well, which brings high resolution on every imaging point. We intend to develop the ultrasound imaging platform with some of the SAFT schemes implemented. The choice of the scheme must be preceded by examination of the imaging quality parameters. The results of the comparison of different SAFT schemes with conventional beamforming are presented in the paper. The results indicate that SAFT schemes can work better than BFR scheme. For example, for some point located near the center of the image the full width at half maximum (FWHM) was equal approximately 0.5, 0.4 and 0.3 mm, while the contrast-to-noise ratio (CNR) was equal 18, 19, and 19 dB for BFR, STA and PWI respectively.
Keywords :
array signal processing; biomedical ultrasonics; ultrasonic focusing; ultrasonic imaging; CNR; FWHM image; PWI; SAFT; classical BFR; classical beamforming; contrast-to-noise ratio; dynamic focusing; experimental results; full width at half maximum image; imaging quality parameter; medical ultrasound system; multielement probe; plane wave imaging; radar technique; synthetic aperture focusing technique; ultrasound imaging platform; Apertures; Array signal processing; Focusing; Signal to noise ratio; Ultrasonic imaging; Wires; Plane Wave Imaging; SAFT; Ultrasonic imaging;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Ultrasonics Symposium (IUS), 2012 IEEE International
Conference_Location :
Dresden
ISSN :
1948-5719
Print_ISBN :
978-1-4673-4561-3
Type :
conf
DOI :
10.1109/ULTSYM.2012.0320
Filename :
6562455
Link To Document :
بازگشت