    In this paper, we address two key issues of a thermoelectric reference standard-based microcalorimeter for RF and microwave power standards. First, we introduce a design guideline of thermoelectric reference standards and present some examples implemented by customizing 3.5-mm and 2.4-mm coaxial power sensors which are commercially available. Second, we propose a simplified measurement system employing a wideband signal source for the thermoelectric reference standard-based microcalorimeters.
  • Keywords
    calorimeters; microwave measurement; standards; RF; coaxial power sensors; measurement system; microcalorimeters; microwave power standards; size 2.4 mm; size 3.5 mm; thermoelectric reference standards; wideband signal source; Radio frequency; Standards; Thermal sensors; Thermistors; Transmission line measurements; Voltage measurement; microcalorimeter; power measurement; reference standards; thermoelectric power sensor;