DocumentCode
63124
Title
Development and Characterization of a New Cylindrical Ionization Chamber for Dosimetry of
Beams
Author
Neves, Lucio P. ; Perini, Ana P. ; Caldas, L.V.E.
Author_Institution
Inst. de Pesquisas Energeticas e Nucl., Comissao Nac. de Energia Nucl., Sao Paulo, Brazil
Volume
60
Issue
2
fYear
2013
fDate
Apr-13
Firstpage
712
Lastpage
715
Abstract
A cavity ionization chamber, made of PVC coated with graphite, was developed and tested to verify its applicability as a reference dosimeter for 60Co beams. In order to characterize this new ionization chamber, several tests were made, following international recommendations, as saturation, ion collection efficiency, polarity effect, short- and medium-term stabilities, linearity of response, angular dependence and leakage current. All results obtained were within those recommended internationally, showing that this prototype may be used as a dosimetric standard system at the Calibration Laboratory of IPEN.
Keywords
dosimetry; ionisation chambers; radiation therapy; 60Co beam dosimetry; IPEN Calibration Laboratory; cavity ionization chamber; cylindrical ionization chamber; dosimetric standard system; international recommendations; ion collection efήciency; leakage current; medium-term stability; polarity effect; radiation dosimetry; radiotherapy; short-term stability; Dosimetry; Ionization chambers; Laboratories; Materials; Standards; X-rays; Gamma beam dosimetry; ionization chamber; radiotherapy;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.2012.2219319
Filename
6340372