Multi-layered fabric electrode for movement artifact reduction in capacitive ECG measurement
Author
Fukuyama, Yasuhiro ; Suzuki, Ryo ; Takayama, S. ; Ueno, Atsushi
Author_Institution
Nissan Res. Center, Nissan Motor Co., Ltd., Kanagawa, Japan
fYear
2013
fDate
3-7 July 2013
Firstpage
555
Lastpage
558
Abstract
An electrode configuration composed of multi-layered fabric electrodes were examined and compared with that composed of non-layered fabric electrodes in terms of movement artifact reduction when using the electrodes in capacitive electrocardiogram measurement from human buttocks in a vibrating environment. Experiments performed with six participants revealed that the multi-layered configuration composed of sensing electrode, driven shield and ground layers reduced amplitude of movement artifact significantly than the non-layered configuration.
Keywords
biomedical electrodes; electrocardiography; capacitive ECG measurement; capacitive electrocardiogram measurement; driven shield layer; electrode configuration; ground layer; human buttock; movement artifact amplitude reduction; movement artifact reduction; multilayered configuration; multilayered fabric electrode; nonlayered fabric electrode; sensing electrode layer; vibrating environment; Atmospheric measurements; Electrocardiography; Electrodes; Fabrics; Force; Particle measurements; Sensors;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society (EMBC), 2013 35th Annual International Conference of the IEEE
Conference_Location
Osaka
ISSN
1557-170X
Type
conf
DOI
10.1109/EMBC.2013.6609560
Filename
6609560
Link To Document