Title :
Fast grating-based X-ray phase-contrast tomosynthesis
Author :
Yan Xi ; Jun Zhao
Author_Institution :
Sch. of Biomed. Eng., Shanghai Jiao Tong Univ., Shanghai, China
Abstract :
As an imaging technique with low radiation dose and improved contrast, digital x-ray tomosynthesis is widely used in clinical diagnoses. Based on the superior capability of x-ray phase-contrast imaging (PCI) techniques for imaging low density materials, the combination of X-ray tomosynthesis and PCI can potentially provide higher efficiency in the detection of soft tissues. The goal of this work was to develop a fast imaging method for phase-contrast tomosynthesis, called fast grating-based phase-contrast tomosynthesis (GPC-Tomo), which integrates tomosynthesis with a grating-based PCI technique. Following the interlaced phase-stepping (PS) data collection method, which is much faster than conventional PS method, we propose a novel image reconstruction method called inner-focusing (IF) reconstruction for the fast GPC-Tomo. The proposed IF reconstruction method was validated by real experiments and the results suggested its effectiveness in achieving a fast GPC-Tomo.
Keywords :
biological tissues; computerised tomography; diagnostic radiography; image reconstruction; medical image processing; GPC-Tomo; X-ray phase-contrast imaging techniques; clinical diagnoses; digital X-ray tomosynthesis; fast grating-based X-ray phase-contrast tomosynthesis; fast imaging method; grating-based PCI technique; image reconstruction method; improved contrast; inner-focusing reconstruction; interlaced phase-stepping data collection method; low density materials; low radiation dose; soft tissues; Data collection; Gratings; Image reconstruction; Materials; Reconstruction algorithms; X-ray imaging;
Conference_Titel :
Engineering in Medicine and Biology Society (EMBC), 2013 35th Annual International Conference of the IEEE
Conference_Location :
Osaka
DOI :
10.1109/EMBC.2013.6610002