• DocumentCode
    636575
  • Title

    Control of rapid hypothermia induction by total liquid ventilation : Preliminary results

  • Author

    Nadeau, Mathieu ; Micheau, P. ; Robert, Raymond ; Avoine, Olivier ; Tissier, Renaud ; Germim, Pamela Samanta ; Walti, Herve

  • Author_Institution
    Mech. Eng. Dept., Univ. de Sherbrooke, Sherbrooke, QC, Canada
  • fYear
    2013
  • fDate
    3-7 July 2013
  • Firstpage
    3757
  • Lastpage
    3760
  • Abstract
    Mild therapeutic hypothermia (MTH) consists in cooling the body temperature of a patient to between 32 and 34°C. This technique helps to preserve tissues and neurological functions in multi-organ failure by preventing ischemic injury. Total liquid ventilation (TLV) ensures gas exchange in the lungs with a liquid, typically perfluorocarbon (PFC). A liquid ventilator is responsible for ensuring cyclic renewal of tidal volume of oxygenated and temperature-controlled PFC. Hence, TLV using the lung as a heat exchanger and PFC as a heat carrier allows ultra fast cooling of the whole body which can help improve outcome after ischemic injuries. The present study was aimed to evaluate the control performance and safety of automated ultrarapid MTH induction by TLV. Experimentation was conducted using the Inolivent-5.0 liquid ventilator equipped with a PFC treatment unit that allows PFC cooling and heating from the flow of energy carrier water inside a double wall installed on an oxygenator. A water circulating bath is used to manage water temperature. A feedback controller was developed to modulate inspired PFC temperature and control body temperature. Such a controller is important since, with MTH induction, heart temperature should not reach 28°C because of a high risk of fibrillation. The in vivo experimental protocol was conducted on a male newborn lamb of 4.7 kg which, after anesthetization, was submitted to conventional gas ventilation and instrumented with temperature sensors at the femoral artery, oesophagus, right ear drum and rectum. After stabilization, TLV was initiated with fast automated MTH induction to 33.5°C until stabilization of all temperatures. MTH could be reached safely in 3 minutes at the femoral artery, in 3.6 minutes at the esophagus, in 7.7 minutes at the eardrum and in 15 minutes at the rectum. All temperatures were stable at 33.5 ± 0.5°C within 15 minutes. The present results reveal that ultra-fast MTH indu- tion by TLV with Inolivent-5.0 is safe for the heart while maintaining esophageal and arterial temperature over 32.6°C.
  • Keywords
    biological tissues; biomedical equipment; blood vessels; cardiology; diseases; hyperthermia; injuries; lung; neurophysiology; patient treatment; temperature sensors; MTH induction; PFC cooling; PFC heating; PFC temperature; PFC treatment unit; anesthetization; arterial temperature; automated ultrarapid MTH induction; body temperature; control body temperature; control performance; conventional gas ventilation; eardrum; energy carrier water flow; esophageal temperature; esophagus; fast automated MTH induction; feedback controller; femoral artery; gas exchange; heart; heart temperature; heat carrier; heat exchanger; high risk fibrillation; in vivo experimental protocol; inolivent-5.0 liquid ventilator; ischemic injuries; ischemic injury; lung; male newborn lamb; mild therapeutic hypothermia; multiorgan failure; neurological functions; oxygenation; oxygenator; perfluorocarbon; rapid hypothermia induction control; rectum; temperature 32 degC to 34 degC; temperature 33.5 degC; temperature sensors; temperature-controlled PFC; tidal volume; time 15 min; time 3 min; time 3.6 min; time 7.7 min; tissues; total liquid ventilation; ultrafast MTH induction; ultrafast cooling; water circulating bath; water temperature; Heart; Liquids; Lungs; Temperature measurement; Ventilation;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society (EMBC), 2013 35th Annual International Conference of the IEEE
  • Conference_Location
    Osaka
  • ISSN
    1557-170X
  • Type

    conf

  • DOI
    10.1109/EMBC.2013.6610361
  • Filename
    6610361