DocumentCode
636584
Title
Estimation of respiratory signal from thoracic impedance cardiography in low electrical current
Author
Fernando, Jeffry Bonar ; Morikawa, Kota ; Ozawa, Jun
Author_Institution
Adv. Technol. Res. Labs., Panasonic Corp., Kyoto, Japan
fYear
2013
fDate
3-7 July 2013
Firstpage
3829
Lastpage
3832
Abstract
A new method to estimate respiratory signal from thoracic impedance is proposed. To realize battery powered, wearable respiratory monitoring devices, low current impedance measurement techniques are desired. However, under low current conditions, conventional methods to separate cardiac and respiratory signals do not work well as the cardiac signal is much larger than the respiratory signal. In the proposed method, respiratory signal is estimated by calculating an envelope curve from the detected T waves of cardiac component. The results of the experiments show that the accuracy of proposed method is greater than conventional method.
Keywords
bioelectric potentials; electric current; electric impedance; electrocardiography; medical signal processing; patient monitoring; battery powered wearable respiratory monitoring devices; cardiac component; cardiac signal; detected T waves; low-current impedance measurement; low-electrical current; respiratory signal estimation; thoracic impedance cardiography; Accuracy; Electrocardiography; Electrodes; Estimation; Impedance; Impedance measurement; Monitoring;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society (EMBC), 2013 35th Annual International Conference of the IEEE
Conference_Location
Osaka
ISSN
1557-170X
Type
conf
DOI
10.1109/EMBC.2013.6610379
Filename
6610379