DocumentCode
651475
Title
Framework for evaluating EEG signal quality of dry electrode recordings
Author
Tautan, Alexandra-Maria ; Serdijn, Wouter ; Mihajlovic, Vojkan ; Grundlehner, Bernard ; Penders, Julien
Author_Institution
Fac. of Electr. Eng., Delft Univ. of Technol., Delft, Netherlands
fYear
2013
fDate
Oct. 31 2013-Nov. 2 2013
Firstpage
186
Lastpage
189
Abstract
Dry electrodes provide the possibility of moving EEG usage from the research and clinical environment to real life applications. Having a framework for evaluating the performance of dry electrodes would facilitate this process and help EEG system developers to test their designs. This paper describes an evaluation method for dry electrode EEG recordings. The framework includes a setup for synchronous recordings with a parallel dry and gel electrode montage procedure. Several protocols are implemented to evaluate both the time and frequency content of the signal and to compute the setup and settling time. Signal quality was evaluated using signal correlations, SNR and P300 component characteristics. The preliminary data analysis and results show that a comparison between gel and dry electrodes is possible but improvements need to be made to the current evaluation framework.
Keywords
biomedical electrodes; data analysis; electroencephalography; medical signal processing; EEG evaluation method; P300 component characteristics; SNR; current evaluation framework; data analysis; dry electrode EEG recordings; parallel dry-gel electrode montage procedure; signal correlations; signal quality; synchronous recordings; Correlation; Electrodes; Electroencephalography; Hair; Sensors; Signal to noise ratio; Visualization;
fLanguage
English
Publisher
ieee
Conference_Titel
Biomedical Circuits and Systems Conference (BioCAS), 2013 IEEE
Conference_Location
Rotterdam
Type
conf
DOI
10.1109/BioCAS.2013.6679670
Filename
6679670
Link To Document