• DocumentCode
    65495
  • Title

    Minimizing Cytosol Dilution in Whole-Cell Patch-Clamp Experiments

  • Author

    Inayat, Sam ; Pinto, L.H. ; Troy, John B.

  • Author_Institution
    Biomed. Eng. Dept., Northwestern Univ., Evanston, IL, USA
  • Volume
    60
  • Issue
    7
  • fYear
    2013
  • fDate
    Jul-13
  • Firstpage
    2042
  • Lastpage
    2051
  • Abstract
    During a conventional whole-cell patch clamp experiment, diffusible cytosolic ions or molecules absent in the pipette solution can become diluted by a factor of one million or more, leading to diminished current or fluorescent signals. Existing methods to prevent or limit cytosol diffusion include reducing the diameter of the pipette´s orifice, adding cytosolic extract or physiological entities to the pipette solution, and using the perforated patch clamp configuration. The first method introduces measurement error in recorded signals from increased series resistance and the latter two are cumbersome to perform. In addition, most perforated patch configurations, prevent investigators from using test compounds in the pipette solution. We present a method to overcome these limitations by minimizing cytosol dilution using a novel pipette holder. Cell-attached configuration is obtained with the pipette filled with pipette solution. Most of the pipette solution is then replaced with mineral oil so that cytosol dilution can be minimized in whole-cell configuration. To accomplish this requires a suction line and two Ag/AgCl electrodes inside the pipette. Testing our novel pipette holder with Chinese Hamster Ovarian cells, we demonstrate cytosol dilution factors between 76 and 234. For large cells with somas greater than 40 μm, cytosol dilution factors of 10 or less are achievable.
  • Keywords
    biochemistry; biodiffusion; biomedical electrodes; biomedical engineering; cellular biophysics; clamps; fluorescence; molecular biophysics; oils; silver compounds; Ag-AgCl; Ag-AgCl electrodes; Chinese Hamster Ovarian cells; cell-attached configuration; conventional whole-cell patch clamp experiment; cytosol diffusion limiting; cytosol dilution factors; cytosol dilution minimizing; cytosolic extract; diffusible cytosolic ions; fluorescent signal; measurement error; mineral oil; perforated patch clamp configuration; physiological entities; pipette holder; pipette orifice diameter reduction; pipette solution; series resistance; suction line; whole-cell configuration; Assembly; Cells (biology); Electrodes; Electron tubes; Minerals; Polyimides; Wires; Cytosol; electrophysiology; patch clamp; pipette holder; rundown; whole cell; Animals; CHO Cells; Cricetinae; Cricetulus; Cytosol; Diffusion; Equipment Design; Equipment Failure Analysis; Patch-Clamp Techniques;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.2013.2248084
  • Filename
    6468078