DocumentCode :
656125
Title :
Thermoluminescence response of Ge-doped optical fiber dosimeters with different core sizes
Author :
Begum, Momotaz ; Rahman, A. K. M. Mizanur ; Abdul-Rashid, H.A. ; Wan Abdullah, W.S. ; Mohd Noor, N. ; Zulkifli, M.I. ; Muhamad-Yasin, S.Z. ; Mat-Sharif, K.A. ; Tamchek, N. ; Mat Nawi, S.N. ; Wahib, N. ; Amin, Y.M. ; Bradley, David A.
Author_Institution :
Fac. of Eng., Multimedia Univ., Cyberjaya, Malaysia
fYear :
2013
fDate :
28-30 Oct. 2013
Firstpage :
291
Lastpage :
293
Abstract :
Thermoluminescence (TL) properties of five different core diameter of 6 mol% germanium (Ge) doped optical fibers have been investigated for the purpose of TL dosimetry. The optical fiber dosimeter TL properties is compared with commercially available TLD-100 chips (LiF:Mg,Ti). Samples were irradiated using Cobalt-60 standard radiation source ranging from 1Gy to 10 Gy. These fibers show good linear dose response up to 10 Gy. Highest core diameter of Ge doped optical fiber (core 100 μm) provides the best response among all fibers. We observe the larger core fiber show better response than smaller core fiber. The relative sensitivity of 100 μm core optical fiber is 0.26 ± 0.04 with respect to TLD-100 chip.
Keywords :
dosimetry; germanium; optical fibres; thermoluminescence; thermoluminescent dosimeters; Cobalt-60 standard radiation source; TL dosimetry; TLD-100 chip; core sizes; optical fiber dosimeters; radiation absorbed dose 1 Gy to 10 Gy; size 100 mum; thermoluminescence response; Optical fibers; Silicon compounds; Thermoluminescence; TLD-100; doped optical fiber; radiation dosimeter; thermoluminescence (TL);
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Photonics (ICP), 2013 IEEE 4th International Conference on
Conference_Location :
Melaka
Print_ISBN :
978-1-4673-6073-9
Type :
conf
DOI :
10.1109/ICP.2013.6687142
Filename :
6687142
Link To Document :
بازگشت