• DocumentCode
    667283
  • Title

    Comparison of electrical impedance tomography reconstruction techniques applied to IMPETOM system

  • Author

    Santos, Euzeli ; Simini, Franco

  • Author_Institution
    Facultades de Med. e Ing., Univ. de la Republica, Montevideo, Uruguay
  • fYear
    2013
  • fDate
    10-13 Nov. 2013
  • Firstpage
    1
  • Lastpage
    4
  • Abstract
    Electrical Impedance Tomography (EIT) reconstruction can estimate thorax fluid content. Its use in critically ill patients is promising and may prove clinically useful. Boundary voltages (16-electrode frames) were obtained with our 50 kHz-5 mA IMPETOM system. Comparison of the 492-element Newton-Raphson algorithm with EIDORS open source tool (NOSER & GREIT algorithms), applied to a healthy volunteer, suggests that anatomically adjusted 3D models give better results. Nevertheless for phantom imaging an initial uniform image yields more accurate reconstructions. The results help in the selection and implementation of the reconstruction method for systems similar to IMPETOM.
  • Keywords
    Newton-Raphson method; bioelectric potentials; biological organs; biomedical electrodes; electric impedance imaging; image reconstruction; medical image processing; phantoms; 16-electrode frames; 492-element Newton-Raphson algorithm; EIDORS open source tool; GREIT algorithms; IMPETOM system; NOSER algorithms; adjusted 3D models; boundary voltages; critically ill patients; current 5 mA; electrical impedance tomography reconstruction; frequency 50 kHz; initial uniform image; phantom imaging; thorax fluid content; Conductivity; Current measurement; Electrodes; Image reconstruction; Impedance; Tomography; Voltage measurement;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Bioinformatics and Bioengineering (BIBE), 2013 IEEE 13th International Conference on
  • Conference_Location
    Chania
  • Type

    conf

  • DOI
    10.1109/BIBE.2013.6701621
  • Filename
    6701621