    A comparison of heartbeat detectors for the seismocardiogram

  • Author

    Garcia-Gonzalez, M.A. ; Argelagos-Palau, Ariadna ; Fernandez-Chimeno, M. ; Ramos-Castro, J.

  • Author_Institution
    Group of Biomed. & Electron. Instrum. at the Dept. of Electron. Eng., Univ. Politec. de Catalunya (UPC), Barcelona, Spain
  • fYear
    2013
  • fDate
    22-25 Sept. 2013
  • Firstpage
    461
  • Lastpage
    464
  • Abstract
    The study aimed to study the accuracy in RR time series derived from the seismocardiogram when employing different heartbeat detectors in subjects measured in a quiet environment. The ECG and seismocardiogram of 17 healthy volunteers was recorded at a sampling frequency of 5 kHz using a Biopac acquisition system. The seismocardiogram was acquired using a triaxial accelerometer (LIS344ALH, ST Microelectronics). Four detectors of the heartbeat from the seismocardiogram were employed relying either on the Continuous Wavelet Transform or bandpass filtering. The detectors adapt their parameters to the morphology of the signal by estimating mean heart rate and the bandwidth of the signal associated to the heartbeat. For all detectors, the standard deviation of the error in the obtained RR time series is in mean slightly higher than 2 ms and the percentage of obtained RR time intervals that have an error higher than 30 ms is around 3.5%. The seismocardiogram, when measured in a quiet environment, can be used instead of the ECG to obtain reliable RR time series when using proper heartbeat detectors.
  • Keywords
    band-pass filters; biomedical equipment; electrocardiography; medical signal processing; signal sampling; time series; wavelet transforms; Biopac acquisition system; ECG; RR time series; bandpass filtering; continuous wavelet transform; frequency 5 kHz; heartbeat detectors; mean heart rate estimation; sampling frequency; seismocardiogram; signal morphology; standard deviation; triaxial accelerometer; Abstracts; Biomedical measurement; Detectors; Heart beat; Seismic measurements; Time measurement;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Computing in Cardiology Conference (CinC), 2013
  • Conference_Location
    Zaragoza
  • ISSN
    2325-8861
  • Print_ISBN
    978-1-4799-0884-4
  • Type

    conf

  • Filename
    6713413