DocumentCode
678195
Title
Design, modeling and characterization of a 35MHz 1-D CMUT phased array
Author
Xu, Tao ; Tekes, Coskun ; Satir, S. ; Arkan, Evren ; Ghovanloo, Maysam ; Degertekin, F. Levent
Author_Institution
George W. Woodruff Sch. of Mech. Eng., Georgia Inst. of Technol., Atlanta, GA, USA
fYear
2013
fDate
21-25 July 2013
Firstpage
1987
Lastpage
1990
Abstract
High frequency ultrasound arrays have applications ranging from imaging of small animals, skin and eye to intravascular ultrasound (IVUS). We describe an application where a guidewire IVUS system uses high frequency phased arrays with integrated electronics placed directly on the guidewire rather than a catheter for imaging. Multiple arrays provide the full transverse cross section of the artery during percutaneous interventions. We focus on a 1-D CMUT phased array to be used in the guidewire IVUS. CMUTs are particularly suitable for this application with their ease of fabrication and single chip electronics integration. The array frequency is chosen to be around 35-40MHz with 10MHz bandwidth so that the 1-D CMUT phased array with a 300um wide aperture can closely match the lateral resolution of a current 20MHz, 3.5F solid state IVUS array. Here we discuss the initial design, large signal modeling, fabrication and experimental characterization of a 12 element, 300×1000um CMUT array with 25um pitch. The electrical and acoustic characterization results for a CMUT array with 20um square membranes are presented. Modeling results for different size membranes indicating adequate performance for higher bandwidth applications are also discussed.
Keywords
bioelectric phenomena; biomedical electronics; biomedical transducers; biomedical ultrasonics; blood vessels; electric impedance imaging; eye; skin; ultrasonic imaging; ultrasonic transducer arrays; 1D CMUT phased array characterization; 1D CMUT phased array design; 1D CMUT phased array modeling; acoustic characterization; artery; bandwidth 10 MHz; bandwidth applications; electrical characterization; eye; frequency 20 MHz; full transverse cross-section; guidewire intravascular ultrasound system; high frequency ultrasound arrays; integrated electronics; lateral resolution; multiple arrays; percutaneous interventions; signal modeling; single chip electronics integration; size 20 mum; size 300 mum; skin; small animal imaging; square membranes; Acoustics; Arrays; Bandwidth; Fabrication; Frequency measurement; Imaging; Ultrasonic imaging; CMUT; High Frequency Arrays; IVUS; Intravascular ultrasound; Phased Array; guidewire;
fLanguage
English
Publisher
ieee
Conference_Titel
Ultrasonics Symposium (IUS), 2013 IEEE International
Conference_Location
Prague
ISSN
1948-5719
Print_ISBN
978-1-4673-5684-8
Type
conf
DOI
10.1109/ULTSYM.2013.0507
Filename
6725287
Link To Document